Exercise: chemometrics II — applications & validation

Draft skeleton — pending expert review

This exercise is a provisional draft written to accompany the Chemometrics II skeleton lecture while a subject-matter contributor is recruited. The concepts drilled here (task framing, validation, RMSEP) are standard and safe, but the exercise has not yet been reviewed by a chemometrics expert and contains no real dataset. Numbers in the worked problems are small illustrative toys, not measured data.

Goal

By the end of this exercise you should be able to:

  • decide whether a spectroscopy problem is a classification or a regression / calibration task,
  • explain why ordinary least squares (lm) is the wrong tool for wide, collinear spectral data,
  • distinguish RMSEC (training), RMSECV (cross-validated), and RMSEP (independent test), and say which one to trust,
  • compute RMSEP by hand from a small table of predictions,
  • read a components-vs-error curve and choose the number of latent components,
  • spot data leakage — the ways a reported error is made falsely optimistic.

This is a concepts and judgement day. The by-hand skill practised is the error calculation and the reasoning about validation, not any spectral algorithm.

Background

A spectrum is a vector of many correlated measurements (absorbance at many wavelengths). Stack \(n\) samples into a wide matrix \(X\) (\(n \times p\), often \(p \gg n\)) and let \(\mathbf{y}\) be the property or class we want to predict.

Two tasks. Regression / calibration predicts a continuous \(y\) (a concentration); classification predicts a category (a group label).

Why not lm. With \(p \gg n\) and highly collinear columns, \(X^\top X\) is not invertible and an ordinary least squares fit is unstable. Chemometrics instead projects \(X\) onto a few latent components first (PCR, PLS) and models in that low-dimensional space — the machinery from PCA and The OPLS model.

Validation vocabulary. The root-mean-square error of prediction is

\[\mathrm{RMSEP} = \sqrt{\frac{1}{n}\sum_{i=1}^{n}\left(\hat{y}_i - y_i\right)^2},\]

in the same units as \(y\). The same formula gives different names depending on which samples it is computed on:

  • RMSEC — on the training samples (used to fit). Optimistic.
  • RMSECV — on cross-validated held-out predictions (used to tune, e.g. choose the number of components).
  • RMSEP — on an independent test set, touched once. The number you report.

Bias is the mean signed error \(\frac{1}{n}\sum(\hat{y}_i - y_i)\): a systematic offset, ideally near zero.

The cardinal rule. Anything fit on the data — preprocessing, scaling, component count, variable selection — must be decided without the test set, or the reported error is too rosy.

By hand

B1. For each scenario, say whether it is regression (calibration) or classification, and name the target \(y\): (a) predict the protein percentage of a grain sample from its NIR spectrum; (b) decide whether a Raman spectrum comes from a tumour or from healthy tissue; (c) estimate the glucose concentration of a fermentation broth from IR; (d) identify which of four bacterial species an isolate is, from its spectrum.

B2. Explain in one or two sentences why you cannot simply fit lm(y ~ .) on a matrix of 200 wavelengths measured on 30 samples. What quantity fails to exist, and what do PCR / PLS do instead?

B3. A calibration model predicts analyte concentration (units: mg/mL) on a held-out test set of five samples:

sample measured \(y_i\) predicted \(\hat{y}_i\)
1 10.0 10.5
2 12.0 11.0
3 8.0 8.0
4 14.0 15.0
5 11.0 10.5
  1. Compute the residuals \(\hat{y}_i - y_i\).
  2. Compute the RMSEP by hand (use the formula above).
  3. Compute the bias. Is the model systematically high or low?
  4. Since this is the independent test set, which of RMSEC / RMSECV / RMSEP does your number deserve to be called?

B4. The same model was also evaluated on the training samples and gave an error of \(0.2\) mg/mL, whereas your test-set error in B3 is larger. Which number should be reported to a collaborator, and why is the training number smaller?

B5. A PLS calibration is fit with increasing numbers of latent components. The cross-validated error (RMSECV) is:

components 1 2 3 4 5 6 7
RMSECV 3.1 1.8 1.0 0.9 0.92 1.4 2.6
  1. How many components would you choose, and why?
  2. What is happening to the model at 6–7 components?
  3. The training error (RMSEC) at 7 components is \(0.1\) — lower than at 4. Does that mean 7 components is better? Explain.

B6. (leakage hunt). For each workflow, say whether the reported test error is trustworthy or optimistically biased, and why: (a) mean-centre and scale using the mean/SD of the whole dataset, then split into train/test; (b) choose the number of components by picking whichever gives the lowest error on the test set; (c) select the 20 “best” wavelengths using all samples, then run cross-validation on just those 20; (d) fit preprocessing and model on the training set, then apply them unchanged to a test set held out from the start.

B7. (classification metric). A spectral classifier is tested on 100 samples: 90 controls and 10 cases. It labels everything “control”. What is its overall accuracy, and why is that number misleading here? What would you report instead?

In R

Static illustrations only. No packages are installed at render time, so nothing below is executed — read it as a template.

RMSEP from a vector of predictions

y_meas <- c(10.0, 12.0,  8.0, 14.0, 11.0)   # measured
y_pred <- c(10.5, 11.0,  8.0, 15.0, 10.5)   # model predictions

resid <- y_pred - y_meas
rmsep <- sqrt(mean(resid^2))                # same units as y
bias  <- mean(resid)

rmsep   # root-mean-square error of prediction
bias    # systematic offset

Choosing components from a cross-validation curve

ncomp  <- 1:7
rmsecv <- c(3.1, 1.8, 1.0, 0.9, 0.92, 1.4, 2.6)

best <- which.min(rmsecv)   # index of the minimum
best                        # -> 4  (fewest error)
plot(ncomp, rmsecv, type = "b",
     xlab = "number of components", ylab = "RMSECV")

A PLS calibration with built-in cross-validation (sketch)

# Illustrative only: the `pls` package provides plsr().
library(pls)

fit <- plsr(y ~ ., data = train, ncomp = 10, validation = "CV")
validationplot(fit, val.type = "RMSEP")   # inspect the curve
summary(fit)

# predict the untouched test set with the chosen number of components
yhat  <- predict(fit, newdata = test, ncomp = 4)
rmsep <- sqrt(mean((yhat - test$y)^2))
rmsep

Tasks

R1. Reproduce B3 with the RMSEP block above. Confirm your hand-computed RMSEP and bias match the R output.

R2. Enter the B5 table into the components block. Confirm which.min(rmsecv) gives the component count you chose by hand, and describe the shape of the plotted curve in one sentence.

R3. Write two lines of R that compute RMSEP for a second candidate model whose predictions are y_pred2 <- c(9.5, 12.5, 8.5, 13.5, 11.5) against the same y_meas. Which of the two models has the lower RMSEP?

Hints

Ask: is the answer a number on a continuous scale (a concentration, a percentage) or a label from a fixed set of groups (tumour/healthy, one of four species)? A number → regression/calibration; a label → classification.

Ordinary least squares needs to invert \(X^\top X\). With more predictors than samples (\(p > n\)) and collinear columns, that matrix is singular — no unique inverse, so no stable coefficient vector (recall the “no unique solution” case from the Linear systems unit). PCR/PLS sidestep it by regressing on a few latent components instead of the raw wavelengths.

Work column by column: (1) residual \(\hat y_i - y_i\) for each sample; (2) square each residual; (3) average the squares (divide by \(n\), here 5); (4) take the square root. Keep units of \(y\) throughout. See sqrt and mean for the R version: sqrt(mean((y_pred - y_meas)^2)).

Bias is the signed average error mean(y_pred - y_meas) — positive means predictions run high. RMSEP squares first, so it measures magnitude regardless of sign. A model can have near-zero bias but a large RMSEP (errors cancel in the mean but not in the squares).

Training error falls forever as you add components; cross-validated error falls, bottoms out, then rises as the model starts fitting noise. Choose the component count at (or just before) the minimum of the cross-validated curve — not the training curve. When two nearby counts are within noise of each other, prefer the simpler (fewer-component) model. which.min finds the minimum’s index.

Ask of every step: did this use any information from the test set? Preprocessing statistics, component count, and variable selection are all fitted quantities. If any of them saw the test samples (or the whole dataset before splitting), the test error is optimistic. Only a set held out before any fitting, and touched once, gives an honest number.

If 90 of 100 samples are controls, always guessing “control” scores 90% accuracy while catching zero cases. With imbalanced groups, report sensitivity and specificity (or a confusion matrix) rather than overall accuracy.

Solutions

B1. (a) Regression / calibration — \(y\) = protein percentage (continuous). (b) Classification — \(y\) = tumour vs healthy (two groups). (c) Regression / calibration — \(y\) = glucose concentration (continuous). (d) Classification — \(y\) = species identity (one of four groups).

B2. Ordinary least squares requires inverting \(X^\top X\). With 200 wavelengths and only 30 samples (\(p \gg n\)) and strongly correlated columns, that matrix is singular / non-invertible, so there is no unique, stable coefficient vector — the fit is undetermined and wildly sensitive to noise. PCR and PLS instead compress \(X\) into a few latent components (linear combinations of wavelengths) and regress \(y\) on those, which is well-posed. PLS chooses components that are relevant to \(y\); PCR uses the largest-variance components of \(X\) regardless of \(y\).

B3. (a) Residuals \(\hat y_i - y_i\): \(10.5-10.0 = +0.5\); \(11.0-12.0 = -1.0\); \(8.0-8.0 = 0.0\); \(15.0-14.0 = +1.0\); \(10.5-11.0 = -0.5\). (b) Square them: \(0.25,\ 1.00,\ 0.00,\ 1.00,\ 0.25\). Sum \(= 2.50\). Mean \(= 2.50/5 = 0.50\). RMSEP \(= \sqrt{0.50} \approx \mathbf{0.71}\) mg/mL. (c) Bias \(= \frac{1}{5}(0.5 - 1.0 + 0.0 + 1.0 - 0.5) = \frac{0}{5} = \mathbf{0}\) mg/mL — no systematic offset (the errors happen to cancel; note RMSEP is still \(0.71\)). (d) It is computed on the independent test set, so it is honestly called RMSEP.

B4. Report the test-set number (\(\approx 0.71\) mg/mL, the RMSEP). The training error (\(0.2\)) is smaller because the model was fit to those very samples — it has partly memorised them, so its error on them is optimistic and does not reflect performance on new samples. Only the independent test set estimates future performance.

B5. (a) Choose 4 components (or arguably 3): RMSECV is minimised at 4 (\(0.9\)), and 3 (\(1.0\)) is nearly as good — if you favour simplicity, 3 is defensible, but 4 is the strict minimum. (b) At 6–7 components the cross-validated error rises sharply (\(1.4\), then \(2.6\)): the model is overfitting, fitting noise in the training spectra that does not generalise. (c) No. A lower training error (RMSEC) at 7 components just confirms the model is fitting the training data more closely, including its noise. The cross-validated error, which reflects new samples, is worse at 7 than at 4. Trust the cross-validated curve, not the training curve.

B6. (a) Optimistically biased. The centering/scaling statistics were computed using the test samples, leaking information from test into the fitted preprocessing. (b) Optimistically biased. Tuning the component count on the test set makes the test set part of model selection — it is no longer independent. Choose components by cross-validation within the training data. (c) Optimistically biased. Selecting wavelengths using all samples (including those later used to estimate error) lets the selection peek at the validation data; selection must happen inside the cross-validation loop. (d) Trustworthy. Preprocessing and model were fit on the training set only and applied unchanged to a test set held out from the start — no leakage.

B7. Accuracy \(= 90/100 = \mathbf{90\%}\), but it is misleading because the classifier catches none of the 10 cases (sensitivity \(= 0\)) — the high number comes entirely from the majority class. Report sensitivity and specificity (or the full confusion matrix); with imbalanced groups, overall accuracy hides total failure on the minority class.

R1.

y_meas <- c(10.0, 12.0,  8.0, 14.0, 11.0)
y_pred <- c(10.5, 11.0,  8.0, 15.0, 10.5)
sqrt(mean((y_pred - y_meas)^2))   # 0.7071068  -> matches ~0.71
mean(y_pred - y_meas)             # 0          -> bias is zero

Matches the hand answers: RMSEP \(\approx 0.71\), bias \(= 0\).

R2.

ncomp  <- 1:7
rmsecv <- c(3.1, 1.8, 1.0, 0.9, 0.92, 1.4, 2.6)
which.min(rmsecv)    # 4  -> minimum at 4 components

which.min returns 4, matching the by-hand choice. The curve falls steeply from 1 to 3–4, flattens at its minimum, then rises from 5 onward — the classic underfit-then-overfit U-shape.

R3.

y_meas  <- c(10.0, 12.0,  8.0, 14.0, 11.0)
y_pred2 <- c( 9.5, 12.5,  8.5, 13.5, 11.5)
sqrt(mean((y_pred2 - y_meas)^2))   # 0.5  -> lower than 0.71

The residuals are all \(\pm 0.5\), so RMSEP \(= \sqrt{\text{mean}(0.25)} = 0.5\) mg/mL — lower than the first model’s \(0.71\), so the second model predicts this test set more accurately.

Where this leads

  • PCA and The OPLS model supply the latent-variable machinery (scores, components) that every method here builds on; this unit is about applying and validating them.
  • Chemometrics I covers the spectra themselves and preprocessing in more detail; revisit it for the signal-processing steps named only briefly here.
  • The validation discipline — hold out an independent test set, tune by cross-validation, report error in interpretable units — is not specific to spectra. It carries into every later data-analysis unit, including the single-cell applications.
Warning

Reminder: this is a draft skeleton. A chemometrics expert should replace the toy tables with a real spectral dataset and add a hands-on PLS calibration once the unit is staffed.